5.3 Measurement specifications
141
Voltage Measurement Specifications (U: Displayed as V on panel display)
Current Measurement Specifications (I: Displayed as A on panel display)
Maximum effective peak
voltage
±600% of each voltage range
However, for 300 V, 600 V, and 1,000 V ranges, ±1,500 Vpeak
Maximum effective peak
current
±600% of each current range
However, for 20 A range and 50 A range, ±100 Apeak
Guaranteed accuracy period
1 year
Conditions of guaranteed
accuracy
Temperature and humidity for guaranteed accuracy: 23°C ±5°C, 80% RH or less
Warm-up time: 30 minutes
Input: Sine wave input, power factor of 1, voltage to earth of 0 V, after zero-adjustment;
within range in which the fundamental wave satisfies synchronization source
conditions
Temperature coefficient
±0.03% f.s. per °C or less
Power factor effects
±0.1% f.s. or less (45 to 66 Hz, at power factor = 0)
Internal circuitry voltage/current phase difference: ±0.0573°
Effect of common mode
voltage
±0.02% f.s. or less (600 V, 50/60 Hz, applied between input terminals and enclosure)
Magnetic field interference
400 A/m, DC and 50/60 Hz magnetic field
Voltage
±1.5% f.s. or less
Current
±1.5% f.s. or ±10 mA, whichever is greater, or less
Active power ±3.0% f.s. or (voltage influence quantity) × (±10 mA),
whichever is greater, or less
Magnetization effect
±10 mA equivalent or less (after inputting 100 A DC to the current direct input terminals)
Adjacent channel input effect
±10 mA equivalent or less (when inputting 50 A to adjacent channel)
Measurement types
Rectifiers: AC+DC, DC, AC, FND, AC+DC Umn
Effective measuring range
1% to 130% of range; however, up to ±1,500 V peak value and 1,000 V RMS value
Display range
0.5% to 140% of range (zero-suppression when less than 0.5%)
Polarity
Displayed when using DC rectifier
Over-range display
Displays [o.r] (over-range) when input exceeds ±140% of the range or ±1,060.5 V.
Peak exceeded warning
The PEAK OVER U (U1 to U3) lamp will light up for any channel for which the input volt-
age peak value exceeds ±1,500 V or ±600% of the range.
Measurement types
Rectifiers: AC+DC, DC, AC, FND, AC+DC Umn
Effective measuring range
1% to 130% of range
Display range
0.5% to 140% of range (zero-suppression when less than 0.5%)
Polarity
Displayed when using DC rectifier
Over-range display
Displays [o.r] when input exceeds ±140% of the range.
Peak exceeded warning
The PEAK OVER I lamp (I1 to I3) will light up for any channel for which the input current
peak value exceeds ±100 A or ±600% of the range.
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......