4.2 LAN Settings and Connection (Before Using Command Communications)
128
Making 1:1 Connections Between the Instrument and a PC
This completes the process of connecting the instrument and PC.
1
Connect the 9642 LAN Cable to the supplied
crossover adapter.
2
Connect the crossover adapter to the 100BASE-
TX connector on the instrument.
3
Connect the 9642 LAN Cable to the 100BASE-TX
connector on the PC.
3
2
Model 9642 LAN Cable
1
Connection cable: Use one of the following.
• 100BASE-TX crossover cable (maximum length 100 m)
• 100BASE-TX straight-through cable with crossover adapter
(maximum length 100 m)
• 9642 LAN cable (option, supplied with crossover adapter)
Connecting with the 9642 LAN Cable
and crossover adapter (supplied)
100BASE-TX connector on instrument
Crossover adapter
Connect to PC
Instrument
PC
100BASE-TX
cross cable
You can monitor and control the instrument from a PC
by connecting the instrument to the PC with LAN cable
(100BASE-TX cable)
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......