5.3 Measurement specifications
139
Basic Measurement Specifications
5.3
Measurement specifications
Measurement method
Simultaneous voltage and current digital sampling, zero-cross simultaneous calculation
Sampling frequency
Approx. 700 kHz
A/D converter resolution
16-bit
Frequency bands
DC, 0.1 Hz to 100 kHz (Reference value of 0.1 Hz
≤
f < 10 Hz)
Synchronization sources
(SYNC)
U1, U2, U3, I1, I2, I3, DC (fixed at 200 ms)
Can be set separately for each wiring mode.
When the source input AC measured value is less than 1% f.s., operation and accuracy
are not defined.
When synchronization cannot be detected, the SYNC ERROR lamp will light up.
Operation and accuracy are not defined at that time.
When the synchronization source is DC, accuracy is not defined for input with a cycle
that is not divisible by 200 ms.
Synchronization timeout setting: 0.1 sec. / 1 sec. / 10 sec. (linked to measurement lower
limit frequency setting)
Measurement items
Voltage, current, active power, apparent power, reactive power, power factor, phase an-
gle, frequency, current integration, active power integration, integration time, voltage
waveform peak value, current waveform peak value, efficiency, voltage crest factor, cur-
rent crest factor, time average current, time average active power, voltage ripple rate,
current ripple rate
Harmonic parameters
Harmonic voltage RMS value, harmonic current RMS value, harmonic active power, to-
tal harmonic voltage distortion, total harmonic current distortion, fundamental wave volt-
age, fundamental wave current, fundamental wave active power, fundamental wave
apparent power, fundamental wave reactive power, fundamental wave power factor
(displacement power factor), fundamental wave voltage current phase difference, inter-
channel voltage fundamental wave phase difference, inter-channel current fundamental
wave phase difference, harmonic voltage content percentage, harmonic current content
percentage, harmonic active power content percentage
The following parameters can be downloaded as data with communications but not dis-
played:
Harmonic voltage phase angle, harmonic current phase angle, harmonic voltage cur-
rent phase difference
Rectifiers
AC+DC
AC+DC measurement Display of true RMS values for both voltage and
current
AC+DC Umn AC+DC measurement Display of average value rectified RMS convert-
ed values for voltage and true RMS values for
current
DC
DC measurement
Display of simple averages for both voltage and
current
Display of values calculated by (voltage DC val-
ue) × (current DC value) for active power
AC
AC measurement
Display of values calculated by
for both
voltage and current
Display of values calculated by
(AC+DC value) - (DC value) for active power
FND
Extraction and display of the fundamental wave component from har-
monic measurement
Zero-Crossing Filter
100 Hz/ 500 Hz/ 5 kHz/ 200 kHz (linked to frequency measurement range)
100 Hz:
0.1 Hz to 100 Hz
500 Hz:
0.1 Hz to 500 Hz
5 kHz:
0.1 Hz to 5 kHz
200 kHz: 0.1 Hz to 200 kHz
The zero-cross filter lower limit setting is linked to the synchronization timeout setting.
10 Hz/ 1 Hz/ 0.1 Hz
(AC+DC value)
2
- (DC value)
2
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......