Every time
key is pressed, the indi-
cation changes in the order: resistance
→
current
→
DAR 1 min/15s
→
DAR 1 min/30s
→
PI
→
resistance
→
current
→
etc.
❖
PI/DAR
→
See 4.2 "Displaying PI and DAR"
(page 83).
3.2.6
Insulation Resistance Measurement
Basis
When a high DC voltage is applied to an
object under test, a leakage current flows.
The insulation resistance tester measures
the applied voltage V and the combined
leakage current I and then calculates the
insulation resistance R.
Calculation formula R = V/I
I
C
and I
A
gradually decrease after the volt-
age is applied.
Holding data after measurement
If the indication is unstable, press the
key. The
average of the measurements is shown.
[
< 1.00 nA
] means "below 1.00 nA."
Applied
voltage
V
Insulation
tester
Capacity
C
Insulation
resistance
R
Steady-state
leakage
current
I
R
Charge
current
I
C
Absorption
current
I
A
Insulating material
C
A
Combined leakage current I
A
Current
detector
R
A