3.1 Pre-Operation Inspection
54
Inspection
Procedure
1.
Clip the resistor to the red and black test
leads connected to the tester.
2.
Also, clip the resistor to the test lead of
the high-voltage meter.
3.
Set the test voltage of the tester to
[
5.00 kV
].
❖
See 3.2Measuring Insulation Resistance, Proce-
dure 5. (page 60) to 8. (page 60).
4.
Hold down
key for more than
one second to start insulation resistance
measurement.
5.
Check to see if the reading of the high-volt-
age meter is somewhere between 5 kV and
5.5 kV.
6.
Check to see if the voltage reading of the
tester is somewhere between 5 kV and
5.5 kV.