5.1 Recording Measurement Data
104
Procedure
1.
Measure insulation resistance or tem-
perature and stop measurement. (Tem-
perature and humidity can be entered by
key operation.)
2.
Press the
key.
[
MEMO No.
] lights up and the No. of the
last stored No. will blink.
3.
Choose data No. using the
key.
Press the
key to display a data
number of another group.
Example
: . . .
↔
A0
↔
b0
↔
C0
↔
. . .
If the
and
keys are held down
simultaneously, the lowest number
among the available data numbers
appears.
Temperature only or temperature and
humidity may be stored as a manual record.
The tester, however, has to be in the
standard measurement mode (both [
STEP
]
and [
TC
] off). They cannot be recorded in the
step voltage test mode ([
STEP
] on) or in the
temperature correction mode ([
TC
] on).
❖
Change voltage setting
→
3.2.1 Procedure 5. to 8. (page 60)
❖
Exit temperature correction mode
→
4.3.2 "Exiting Temperature Correction Mode"
(page 92)
❖
Enter temperature/humidity by key operation.
→
6.3 "Entering Temperature and Humidity Mea-
sured with External Thermometer and Hygrome-
ter" (page 129)