Installation
Procedure
Tools: Phillips screwdriver
1.
Turn off power and disconnect the test leads,
AC adapter and USB Cable from the tester.
❖
See 2.2 "Turning Power On and Off" (page 44).
2.
Loosen the set screw on the rear of the
tester and remove the battery cover.
• If the battery pack is not used for an
extended period of time, remove it from the
tester and store at a temperature between
-20 to 30°C, to prevent deterioration.
• Charge the battery at least every 2
months. If the battery pack is left for a long
period of time in a low state of charge, its
performance will be degraded.
• When the battery status indicator is low,
charge the battery pack.
• The charge stored in the battery pack
naturally dissipates over time, therefore be
sure to charge the battery pack before use.
If the operating time is extremely short
directly after the battery pack has been
charged, the battery needs to be replaced.
• The life of the battery pack is 500
charging cycles, i.e., about one year.
Set screw
Remove
Battery cover