![Hioki 3455 Instruction Manual Download Page 116](http://html.mh-extra.com/html/hioki/3455/3455_instruction-manual_2137473116.webp)
5.1 Recording Measurement Data
1
2
3
4
5
6
7
8
9
10
11
付録
索引
113
11.
Insulation resistance measurement stops
under one of the three conditions below.
1. Time equal to recording interval x maximum
number of loggings has elapsed.
2. The set time of the timer has elapsed.
3. The
key is pressed.
After measurement has been completed,
the data No. blinks.
Data is not stored in the memory at this
point.
12.
Measure temperature, if necessary. This
may be omitted.
Temperature and humidity measured with
external thermometer and hygrometer
may be entered by key operation.
❖
See 3.4 "Measuring Temperature" (page 76).
❖
See 6.3 "Entering Temperature and Humidity
Measured with External Thermometer and
Hygrometer" (page 129).
• If measurement is stopped before the first recording
interval elapses, no logging records are acquired and
[
MEMO No.
] and the data No. are turned off.
• When the data No. blinks upon completion of mea-
surement, if [
LObAt
] appears due to low battery or if
power is turned off by the auto power off, the data will
be stored in the memory.