2.1 Supplying Power
1
2
3
4
5
6
7
8
9
10
11
付録
索引
41
2.1.4
Charging the Battery Pack
The 9459 BATTERY PACK can be charged
while installed in the tester, using the
optional 9753 AC ADAPTER.
Short charge time: Approx. 3 hours (at
23°C room temperature)
Procedure
1.
Install the battery pack.
❖
See 2.1.2 "Installing the Battery Pack
(Rechargeable nickel-hydrogen battery)" (page
34).
• Carry out battery charging at an ambient
temperature between 10°C and 40°C.
However, the ambient temperature may
influence the charging efficiency. Outside
this range, not only is the charging capacity
reduced, but also there is a possibility of
reduced performance or electrolyte leakage.
• The battery pack cannot be charged when
test leads are connected to the tester.
• The battery pack will be charged regardless
of the battery selector switch position.
• Communication with a PC and temperature
measurement are available during charging.
But, insulation resistance measurement and
voltage measurement are not available.
• Only use the specified battery charger.
• Do not recharge a fully-charged battery
pack. If the battery pack is over-charged,
a deterioration in performance or battery
fluid leakage may result.