7.2 Measurement Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
157
7.2.4
Temperature Measurement
Temperature and humidity range for guar-
anteed accuracy: 23±5°C less than 90%RH
(non-condensation)
Accuracy when using with the 9631-01,
9631-05 TEMPERATURE SENSOR
Measurement Range, Accuracy
When the 9631-05 TEMPERATURE SENSOR is
used, the accuracy is guaranteed within 0.0 to 40.0°C.
Measuring range
Measurement
accuracy
-10.0°C to -0.1°C
±1.5°C
0.0°C to 40.0°C
±1.0°C
40.1°C to 70.0°C
±1.5°C
Temperature
characteristics
Measurement accuracy × 0.5 is added to
the measurement accuracy. (when the
ambient temperature is not 23±5°C)
Response time
Approx. 100 sec. Including the period of
time for the response of the 9631-01,
9631-05 TEMPERATURE SENSORs.
(Reference value: Period of time until
90% of the change in temperature is
reflected in the indication)
Influence of
radioactive RF
electromagnetic
field
±2°C at 3V/m
www.
.com
1.800.561.8187