4.3 Setting the Recording Method [REC&EVENT]
66
When timed recording is enabled, recording start and stop times can be set.
Setting Recording Start and Stop Times
Move to the setting item.
Select the screen.
1
2
Setting Contents
OFF
Start and stop recording by pressing the
START/STOP
key.
However, note that after pressing the
START/STOP
key, and depending on
the specified interval, recording starts only at appropriate times.
ON
Operates as though you pressed the
START/STOP
key at the specified start
and stop times.
Select from the pull-down menu.
3
Accept
Cancel
When OFF (recording start time depends on interval setting)
Example: If the following intervals are selected and you press START/STOP at current time
10:41:22
Interval Setting
10:45:00
11:00:00
Recording does not actually start until the begin-
ning of the next scheduled interval.
[WAITING]
5 min
30 min
When ON
When an interval is specified, recording does not actually start until the appropriate time.
Start and stop times can be set by two methods.
Selects year, month, day, hour and minute
Change a numerical value ([
]: increase the value, [
]:
decrease the value)
Automatic Setting
[Start Time]
[Stop Time]
Manual Setting
Sets to the current time
Sets to one hour from the start time.
Sets one day from the start time.
Sets one week from the start time.
10:41:22
If the start time has already passed when you press the
START/STOP
key,
"Operation Error!" is displayed. During the
[WAITING]
mode, pressing the
START/STOP
key before the start time aborts the recording.
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...