5.6 Recovering From a Long-Term Power Outage
94
If the power supplied to the instrument is interrupted, operation contin-
ues on the (charged) battery pack. However, if the power is not restored
for a long time (four to six hours or more), the instrument shuts off.
However, if memory partitioning is set [ON], and power fails while
recording, recording resumes in the next partition when power is
restored (for example, if power fails while recording in Memory Partition
2, recording resumes in Partition 3).
The following message appears initially when power is restored, and
recording resumes immediately.
(operating at 23°C)
5.6
Recovering From a Long-Term Power
Outage
HIOKI 3197 POWER QUALITY ANALYZER V 1.00
Recording Re-started.
However, if recording in Memory Partition 4 or with memory partitioning
set [OFF] when power is interrupted, recording does not resume when
power is restored (even though "Recording Re-started" still appears)
Internal Operating
State
Partition
Memory No.
Operation when power is
restored
[RECORDING]
[ON]
No. 1, 2, or 3
Recording resumes in the
next memory partition.
No. 4
Recording does not resume.
[OFF]
-
Nominal continuous operating time (battery only)
After full charge, with LCD backlight auto-off (after 5 min.) enabled: Approx. 6 hours
After full charge, with LCD backlight always on : Approx. 4 hours
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...