8.3 Event Specifications
151
8
Timer
Measurement
method
Detect events at preset intervals
Timer event
interval
OFF, 1, 5, 15 or 30 minutes; 1, 2 or 12 hours; or 1 day
Detection criteria Occurrence time
Recording
contents
Event data, event waveform
Event data format Event number, date, time, event type
Manual
Measurement
method
Detect events when keys are pressed (pressing the
[ESC] and [EVENT] keys at the same time)
Recording
contents
Event data, event waveform
Event data format Event number, date, time, event type
Start, Stop
Measurement
method
Recording start and stop operations are detected as
events
Recording
contents
Event data, event waveform,
Event data format Event number, date, time, event type
Event Type
Event Detection
Event detection method
Detection by ORing each event type
Event setting
Events other than manual events, and start/stop can be enabled/
disabled (set on/off)
Event waveform recording
length
Waveform from 20 ms before det 200 ms upon det 30
ms after detection
Event voltage fluctuation
graph length
0.5 s 2.5 s after detection
Inrush current graph
length
0.5 s 29.5 s after detection
Maximum event count
1000
Maximum recording event
number
50 event waveforms
20 event voltage fluctuation graphs
1 inrush current graph
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...