4.2 Selecting the Measurement Method (SYSTEM Screen) [MEASURE]
58
Selecting the Line Frequency (Frequency)
Move to the setting item.
Select the screen.
1
2
Setting Contents
AUTO/ 50Hz/ 60Hz
Accept
Cancel
Select from the pull-down menu.
3
The AUTO setting automatically selects 50 or 60 Hz internally when measurement starts.
Setting the Wiring Method (Wiring)
Move to the setting item.
Select the screen.
1
2
Setting Contents
1P2W
Measure single-phase, 2-wire lines
1P3W
Measure single-phase, 3-wire lines
3P3W2M
Measure three-phase, 3-wire lines (Delta configuration, using the two-
meter method)
(Use to measure three-phase power by measuring current at just two
points.)
3P3W3M
Measure three-phase, 3-wire lines (Delta configuration, using the three-
meter method)
3P4W
Measure three-phase, 4-wire (star configuration)
3P4W2.5E
Measure three-phase, 4-wire, 2.5-element lines (star configuration)
(Use to measure three-phase power by measuring voltage at just two
points.)
Select from the pull-down menu.
3
Accept
Cancel
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...