![Hioki 3197 Instruction Manual Download Page 186](http://html.mh-extra.com/html/hioki/3197/3197_instruction-manual_2137178186.webp)
Appendix 6 Clamp Sensor Combined Accuracy
A
10
Appendix 6
Clamp Sensor Combined Accuracy
Current and Power Accuracy of Clamp Sensor Combinations
Clamp Sensor
3197 R
ange
Combined Accuracy
9657-10
0.5A
±
1.3%rdg.
±
1.2%f.s.
5A
±
1.3%rdg.
±
0.3%f.s.
9675
0.5A
±
1.3%rdg.
±
0.3%f.s.
5A
±
1.3%rdg.
±
0.21%f.s.
9694
5A
±
0.6%rdg.
±
0.22%f.s.
50A
No specified accuracy
9695-02
5A
±
0.6%rdg.
±
0.4%f.s.
50A
±
0.6%rdg.
±
0.22%f.s.
9660
10A
±
0.6%rdg.
±
0.4%f.s.
100A
±
0.6%rdg.
±
0.22%f.s.
9695-03
10A
±
0.6%rdg.
±
0.4%f.s.
100A
±
0.6%rdg.
±
0.22%f.s.
9661
50A
±
0.6%rdg.
±
0.3%f.s.
500A
±
0.6%rdg.
±
0.21%f.s.
9667 500A
50A
±
2.3%rdg.
±
3.2%f.s.
500A
±
2.3%rdg.
±
0.5%f.s.
9669
100A
±
1.3%rdg.
±
0.3%f.s.
1kA
±
1.3%rdg.
±
0.21%f.s.
9667 5kA
500A
±
2.3%rdg.
±
3.2%f.s.
5kA
±
2.3%rdg.
±
0.5%f.s.
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...