6.3 Viewing Anomalous Phenomena (EVENT screen)
113
6
Inrush Current
• Detection Method
Detects using Irms 1/2.
Set the current threshold value.
An event is detected when the current exceeds
the threshold in the positive direction.
• Recording Contents
Event list data, voltage/current waveforms, inrush
current fluctuation graph
• Event Data Format
Event number, occurrence date, occurrence time,
event type, channel and maximum current
• Inrush Current Fluctuation Graph
Fluctuation graph of Irms from about 0.5 seconds
before detection to about 29.5 seconds after
detection
Timer
• Detection Method
A detected event occurs after every set timer
interval.
• Recording Contents
Event list data, voltage/current waveforms
• Event Data Format
Event number, occurrence date, occurrence time,
event type
Manual
• Detection Method
A detected event occurs whenever the
ESC
and
EVENT
keys are pressed simultaneously.
• Recording Contents
Event list data, voltage/current waveforms
• Event Data Format
Event number, occurrence date, occurrence time,
event type
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......
Page 200: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...