6.3 Viewing Anomalous Phenomena (EVENT screen)
117
6
Details of an event selected in the Event List can be viewed in the Event
Detail List. When multiple events have occurred, the details of the events
can be viewed in the Event Detail List. Multiple events occurring within the
same 200-ms period are displayed together as a single event.
Viewing Event Detection Details
[DETAILS]
Select the screen.
Event Type
Channels
Numerical fields are displayed for IN (Onset) events
(Fields are blank for OUT (Recovery) events.)
Time
Voltage
Value
Continuation
Period
The Event Detail List can be scrolled when it
contains more than six items.
Select an event
Summary of Contents for 3197
Page 1: ...3197 POWER QUALITY ANALYZER May 2012 Revised edition 7 3197A981 07 12 05H Instruction Manual ...
Page 2: ......
Page 8: ...Contents vi ...
Page 30: ...1 3 Measurement Flowchart 22 ...
Page 44: ...2 5 Internal Operating Status and Memory Usage 36 ...
Page 86: ...4 6 Initializing the Instrument System Reset 78 ...
Page 134: ...6 4 Viewing Recorded Data REVIEW State 126 ...
Page 176: ...9 4 Disposing of the Instrument 168 ...
Page 199: ......