2-26
888-2523-001
7/26/05
WARNING: Disconnect primary power prior to servicing.
Section 2 Installation / Initial Turn-On
ATLAS Analogue
•
Open test load supply and return line valves.
•
Close cabinet(s) bypass valve(s).
2.11.4 Final Cooling System Fill
!
CAUTION:
THE SYSTEM MUST BE TESTED FOR LEAKS AGAIN ONCE THE REQUIRED 50/50
GLYCOL/WATER MIXTURE HAS FILLED THE ENTIRE SYSTEM.
Fill pump module tank with with equal amounts of coolant and water (see section
B.1 Coolant and Water Recommendations in Appendix B).
Run pumps for several minutes to start refilling the rest of the system. Alternate
pumps A and B.
Check and fill pump module tank with equal amounts of coolant and water as
needed.
Repeat above steps until coolant level has stablized at the Normal Level indication
in the site glass.
NOTE:
Please note that capacities shown include all system components such as trans-
mitter w/contents, pump module, heat exchanger, test load (w/connection), cabi-
net bypass, and multiple cabinet interconnections where used. The values do not
include the main interconnection plumbing lines between the transmitter, pump
Table 2-8 Cooling System Capacities
Transmitter Model
Approximate System Capacity
(less plumbing lines)
DVA2500
48 gallons (182 liters)
DVA5000
48 gallons (182 liters)
DVA10000
53 gallons (200 liters)
DVA20000
70 gallons (265 liters)
DVA30000
83 gallons (314 liters)
Summary of Contents for ATLAS ANALOGUE
Page 4: ...iv 888 2523 001 7 26 05 WARNING Disconnect primary power prior to servicing...
Page 6: ...vi 888 2523 001 7 26 05 WARNING Disconnect primary power prior to servicing...
Page 8: ...viii 888 2523 001 7 26 05 WARNING Disconnect primary power prior to servicing...
Page 10: ...x 888 2523 001 7 26 05 WARNING Disconnect primary power prior to servicing...
Page 14: ...Table of Contents continued 4...