34
Plane6 Screens -
Step 3: Results
When all the data is collected, click on Step 3 icon to bring you to the Step 3: Results. This shows the Moody Method
calculations to get the overall flatness results.
The results are summarized in the Results Table and the plate-grade tolerance, chosen from Step 1, is applied and a red X
or green check-mark is displayed to indicate if the overall flatness is in or out of tolerance.
A graph is also shown giving a snap shot of the flatness data. A more dynamic graphic of the flatness is shown in
Step 4:
Graph
.
The data tables for each line segment are also shown along with the various values needed to calculate the overall flatness.
The
Zero Low Point
column is the corrected flatness value for that line segment. The zero point is the lowest value of all
the sets of data.
Figure 27 - Plane6 Step 3: Results