30
Moody Center Tolerance:
As per
Moody Method
procedure, this is the maximum allowed deviation between
the intersecting diagonal lines after all point adjustments
have been made. If this tolerance is exceeded, it suggests
that measurement errors were made and it suggested that
the data be retaken.
Plate Serial Number:
Enter the plate’s serial number for
the report.
Label Location on Plate:
Location of the manufacturer’s
label applied to the plate. The letters correspond to the
letters on the plate graphic A…H points as shown above.
AB
means the plate is located between A & B on the plate.
Plate Temperature Top/Bottom:
Temperatures for the
top and bottom of the plate (if measured) as the flatness
data was recorded.