Step 4b – Move the T-1297 Base to
the Far End of Straight Edge
a.
Move the T-1297 Base to the
far
end of the straight edge.
b.
Adjust the H (yaw) and V (pitch) axis
adjustments on the L-703S laser to tilt the
beam until the values are less than:
H Axis: 30,000
V Axis: 10,000
This aligns the laser beam to the straight
edge and puts it on the most accurate part
of the PSD sensor.
Step 4c – Move the T-1297 back to Near
Position and Start Recording
Move the T-1297 Target back to the near point
and click
Record
. When positioning the target,
line up the Measuring Plane (front edge of target
base mounting surface) with the ruler dimension
before hitting record. In this case it should be on
the zero (0) point on the ruler
.