38
Step 2 – Place Straight Edge on the Diagonal
Segment
Plane6 preselects the line sequence that you will
measure. This can be overridden but this sequence
was selected to minimize the time it takes take all
the data. To measure the diagonal:
a.
Locate the correct diagonal segment using the
plate label location to orient yourself (see
screen below).
b.
Place the side of the straight edge with the ruler
so it is to the right of the corner locator when
viewed from the Laser End. See image. In
other words, you always want the
scale side
of
the ruler to be touching the locator tool.
c.
Position the end of the straight edge to line up
with the line on the rotating-edge-alignment
tool.
d.
Now the straight edge is located next to the line
segment that you need to measure such that
when the target is put down against the straight
edge, it is directly over the line segment.
Step 3 – Assemble Laser and Laser Mount
Insert the L-703S Laser into the L-703SP-LM
Laser Mounting Fixture and tighten the thumb
screw very tight and make sure to keep the top if
the laser level.
Summary of Contents for L-703SP
Page 1: ......
Page 54: ...50 An example on how to do the Math to get the exact flatness values...
Page 55: ...51 Plane6 Sample Report...