43
Step 7 – Move Laser and Straight Edge to Plate
Outside Edge and Record Data
Move to one of the 2 outside edge segments for the
corner you are on. Plane6 will automatically move
to the next line segment, so pay attention to which
edge to record, always looking at the name plate
locator to orient yourself. Find the straight edge
that best matches the length of that side of the
plate.
Note
– we recommend that the extra length of the
straight edge be no longer than 1/3 of the edge of the
plate. This is to ensure that the straight edge does not
move around when you are taking data. This can
happen because as the straight edge gets longer relative
to the plate edge, it tends to want to fall off the plate.
Again when using the corner edge tool, always put the
scale side touching the tool.
Place the L-703SP-LM Laser Mount on the straight
edge as shown Step 3a above. Place the target at
the near position and you are ready to record.
Step 7b - Record Second Outside Edge of Plate
Locate the second outside edge recommended by
Plane6 and repeat the process in Step 7a.
Summary of Contents for L-703SP
Page 1: ......
Page 54: ...50 An example on how to do the Math to get the exact flatness values...
Page 55: ...51 Plane6 Sample Report...