accessories & options
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 150 of 176
[10.P]
Safety multiple socket
The stable safety multiple socket
[cos100157]
can be used for
the safe connection of a medical h/p/cosmos treadmill with
peripheral devices such as PCs, printer or monitors.
It meets all requirements for multiple sockets for creating
connections of a medical electrical system within the patients
environment.
safety multiple socket [cos100157]
As claimed in the norm EN60601-1 for medical electrical devices, the multiple socked is accessible only with the help of
a tool. With this method an unintentionally made connection of further devices to the system is avoided, which lowers
the risk of an exceedance of the maximum allowed patient-leakage-current of 0.5mA at the complete system.
N
Due to the very high load of treadmills it is never allowed to connect more than 1 treadmill or
other consumers with high load do the safety multiple socket. Use a dedicated line only.
N
The power consumption of the consumers which are connected in addition to the treadmill-
ergometer to the safety multiple socket must be in total lower than 300W.
N
When connecting treadmills to the safety multiple socket, there are no performance-limitations
to expect at low-performance applications (walking, slow running). For high-performance
applications (fast running, sprints, etc.) the use of the safety multiple socket can result in
performance limitations.
N
For high performance applications a 3-phase powered treadmill is recommended.
[10.Q]
Digital (serial) RS232 interface / USB converter
Connecting a running machine to a medical device results in a medical system. Always use IEC
601-1 approved potential isolation components when linking medical devices via interface. Only
authorised trained staff is entitled to perform this connection. This medical system, when
connecting a running machine to a medical device, has to be connected via a potential-balance-
cable with the provided connector bolt and with the provided bearing within the medically used
room.
The interface serves as a data transmitter from the running machine to a PC or other controlling devices. You can also
control the Running-Machine from another controlling device (ECG. o.a.) or with the appropriate Software from a PC, via
interface. Standard for all devices is ONE serial interface COM 1. Optionally there are further serial interfaces COM 2,
COM3 and COM4 available. For further information please contact your local dealer or h/p/cosmos. The interface
protocol h/p/cosmos coscom
©
is available on request or can be found at www.coscom.org. The incorporated protocols
(e.g. for ECG-devices) are listed in the list of options under OPTION 20 / OPTION 21. See the compatibility list of the
RS232 interface in the chapter „Technical data“. More recent computers will have only an USB interface instead of the
RS232 interface. For that case an „USB to RS232 interface-adapter-cable“ is available at h/p/cosmos under order
number cos12769. To control via USB interface the processor must be a Pentium 1.8 GHz or higher.