installation & commissioning
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 37 of 176
[05]
Immediately at first installation an electric safety test / measurment
has to be performed for „Protective Earth Resistance“, „Isolation
Resistance“ and „Leakage Current“ and the values have to be
recorded on a special protocol [cos11690xx] and marked with „first
measured values“. For details please refer to chapter
„maintenance / inspections“. One copy of this protocol
[cos11690xx] remains at the owner’s manual and the original with
the „first measured values“ shall be sent to h/p/cosmos. The
potential equalization cable has to be
connected AFTER the measurements.
[06]
Lubricate the running machine before introduction according to
instruction with 30 ml special silicon oil which is part of the “service
box”.
[07]
Check the tension of the running belt (no slip).
[08]
Adjust the running belt properly before instruction with the
customer.
The allen-key must be pulled out of the screw
immediately after usage due to risk of injury!
[09]
Test the function of the heart rate transmission with the help of a
simulator or a POLAR sender. Search for interferences and
possible reasons.
accomplished
accomplished
accomplished
accomplished
accomplished