operation
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 74 of 176
[5.B18]
Possibilities of interfering in test mode manually
Initiation: Operational mode: test, running-belt is moving.
[5.B18]
Step no. Activity
Keys
Response / Display
[01]
Alteration of speed
a
or
d
press and hold
Speed is being increased / reduced (0-max.) as long as
key get pressed. If SPEED is being reduced down to „0“
= PAUSE-position:
=
indicates: PAUS. Current values will be
sustained. Only valid for the actual program step, no
alteration in the memory. Next program step as in the
profile defined.
[02]
Alteration of elevation
Not activated in every test
f
or
e
press and hold
Elevation is being increased / reduced (0-max.) as long
as key get pressed. Only valid for the actual program
step, no alteration in the memory. Next program step as
in the profile definied.
[03]
Alteration of acceleration level
(only during increase or
decrease of speed)
Press several times
c
Example: For acceleration or deceleration level “3”,
press 3 times the corresponding key and then hold.
Note: Maximum 7 levels, limited by the maximum
acceleration level selected in the Optional Mode /
standard value = 4. During deceleration this function
alternates the degree of deceleration.
[04]
Switch over to the next program
step or one program step back
(not activated in every test).
c
and simultaneous
e
or
f
Jump over to the next program step or jump to the
previous program step
[05]
Switch over to the next mode or
one mode back
(not activated in every test)
c
and simultaneous
a
or
d
Jump over to the next mode or to the previous mode.
This is a perfect way to change from test mode to
manual mode (cool down) without stop of the belt.
For a better documentation we recommend to use a printer linked to the serial interface RS 232 or/and to use an external PC
with the Software h/p/cosmos para graphics
®
.