optional settings
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 80 of 176
[6.A2]
OP27
Minimum acceleration and deceleration
level
The selected minimum level is valid for
all acceleration and deceleration
processes in all modes and profiles.
§
blinking, reports the min. acceleration / deceleration for all
modes and profiles (standard: level 1)
Level of settings: 1 ... 5 but not higher than the value in option 28. Due
to safety reasons, the acceleration / deceleration levels 5, 6 and 7 can
not be selected. Note: The selected acceleration and deceleration level
is NOT valid for control and operation via RS232-interface. In this case
the acceleration and deceleration level is set in option no. 29 resp. set
in the corresponding command of the h/p/cosmos coscom protocol.
OP28
Maximum acceleration and deceleration
level
The selected maximum level is valid for
all acceleration and deceleration
processes in all modes and profiles
§
blinking, reports the max. acceleration / deceleration for all
modes and profiles (standard: level 4)
Note: Due to safety reasons, the acceleration / deceleration levels 5, 6
and 7 can only be selected, if the safety of the subject is provided for
by the use of the respective fall prevention (e. g. by the safety arch).
The maximum acceleration and deceleration level is NOT valid for
control and operation via V24/RS232 interface. In this case the
acceleration and deceleration level is set in option no. 29 resp. set in
the corresponding command of the h/p/cosmos coscom protocol.
OP29
Standard- Acceleration and deceleration
level for RS 232 interface
The selected acceleration and deceleration level is valid for control and
operation via RS232 interface. This option is very helpful, if the
peripheral equipment (e.g. ECG, Ergospirometry, PC) does not offer a
menu for acceleration and deceleration levels
.
§
blinking, reports: 1 ... 5, (Standard: 1) for the acceleration and
deceleration level for all speed commands via RS 232. The maximum
value adjustable depends on the setting of option 28.
Note: If the peripheral equipment sends an acceleration and
deceleration command via the h/p/cosmos coscom protocol, the
selected level in option number 29 is not valid for these speed
commands
.
OP40
Locking and unlocking of the treadmill
OFF = After switching on, the treadmill is completely locked / not
accessible. To unlock the treadmill, press the b, - and START
together at the same time. While locked the display shows "no
ACCESS"
ON = treadmill is unlocked / accessible (standard)
OP41
Locking and unlocking the manual mode
OFF = manual mode is locked / not accessible
ON = manual mode is unlocked / accessible (standard)
OP42
Locking and unlocking the profile mode
OFF = profile mode is locked / not accessible
1 ... 6 = profile mode is unlocked / accessible up to the selected profile
number Standard: 6
Example: selected profile number = 3: The profiles 1-3 can be
selected, the profiles 4 – 6 can not be selected
OP43
Locking and unlocking the cardio mode
OFF = cardio mode is locked / not accessible
ON = cardio mode is unlocked (standard)