optional settings
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 95 of 176
[6.B1]
OP93*
Total time of operation (h)
= Stand-By-time including time of use of
the motor/ running belt
§
and
$
report: hours of operation
%
reports: h
Difference to user option 03: the displayed value can also be changed.
OP94*
Total time of use (h)
= time of use of motor / running belt only
§
and
$
report: hours of operation
%
reports: h
Difference to user option 04: the displayed value can also be changed.
OP95
MCU Serial number
Report and entering of 8-digit serial number with the following format:
YYMM NNNN
YY: 2-digit number of year of production lot
MM: month of production lot
NNNN: 4-digit serial number
OP98
Setting of limitation of usage time
Secret option
OP99
Test-Profile 99 for endurance testing in
the factory / service
With this option a special test profile (profile number 99) for endurance
testing in the factory or in service can be activated. This test profile is
not for normal use, so it has to be de-activated again after using. If the
device is switched off an on again, profile number 99 is deactivated
automatically.
Details of test profile 99:
Elevation from 0 to maximum elevation every 20 min.
Speed 3.0 km/h for 2 hours (for ladder ergometer: 3,0 m/min for 2
hours)
Speed 5.0 km/h for 3 hours (for ladder ergometer: 5,0 m/min for 3
hours)
Speed 8.0 km/h for 8 hours (for ladder ergometer: 8,0 m/min for 8
hours)
STOP after 13 hours.
Adjustable settings:
0
= Test-Profile 99 locked/ de-activated
1
= Test-Profile 99 activated
*Note: It is not allowed to change the total time of operation or the total distance of the running-machine to a value which
does not match to the real value of the device.