technical data
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 130 of 176
h/
p/
cosmos venu
s 200/
75
h/
p/
cosmos venu
s 200/
100 [
a]
h/p/cosmos saturn 250/75
h/p/cosmos saturn 250/100 [b]
h/p/cosmos saturn 250/125 [c]
h/p/cosmos saturn 300/75
h/p/cosmos saturn 300/100 [d]
h/p/cosmos saturn 300/125 r
h/p/cosmos saturn 300/150 r [e]
h/p/cosmos saturn 300/175 r
h/p/cosmos saturn
300/200 r [f]
h/p/cosmos saturn 450/300 rs
External control unit / UserTerminal
(Display & Keyboard)
6 LCD displays / 6 keys / display of mode and unit with LED
Power supply
(standard, read nameplate)
Dedicated line required.
400 Volt AC 3~/N/PE f: 50/60 Hz
12800 VA (min)
model h/p/cosmos saturn 300/200rs: 25600 VA
Fuse (standard, read nameplate)
32 A
64 A
Power consumption (Standard, read
nameplate)
17.0 A
[a] 18.0 A
20.0 A
[b] 21.0 A
[c] 22.0 A
23.0 A
31.3 A
50.0 A
62.0 A
Power input
6760 VA
[a] 7160 VA
7800 VA
[b] 8360 VA
[c] 8750 VA
9200 VA
12500 VA
24000 VA
45000VA
Drive motor capacity
7500 W
7500 W
7500 W
11000 W
22000 W
30000W
Elevation motor capacity
370 W
370 … 2200 W (depends on installation) 2 x 5000W
Safety standards
n
MPG, RL 93/42 EWG,
IEC EN 60601-1 (VDE 0750), IEC EN 60601-1-2, IEC EN 60601-1-4
VDE 0751, EN 957-1:2005, EN 957-6:2001
Leakage current
0.25 mA
on request
Isolation transformer
3 x 2000 VA
3 x 2000 VA
[b] and [c]
3 x 3000 VA
3 x 3000
VA
3 x 4200 VA
3 x 4200 VA
3 x 21000
VA
Safety class / -categories
I / IP20 / B
Mode of operation acc. IEC 601-1
Continuous operation
Field of application / Accuracy
sports SA and medical
[r] = suitable for bicycle and wheel chair application (from running surface 3000 x 1250 mm standard)
Classification according to MDD
II b
Speed / Definition
0.0 ... 40.0 km/h / 0.1 km/h
0.0 ... 11.11 m/s / 0.1 m/s
Allowance* +/-
5
%
Option spezial speed
0.0 ... 30.0 or ... 50.0 or ... 60.0 or ... 80.0 km/h
Angle of inclination**
Definition
-35 % ... +35 %
0.1 %
-27 % ... +27%**
0.1 %
-5% (-25%)
… +25%
Dimension of running surface
length x width in mm
[r] = suitable for bicycle and wheel
chair application
2000 x 750
[a] 2000 x 1000
2500 x 750
[b] 2500 x 1000
[c] 2500 x 1250
3000 x 750
[d] 3000 x
1000
3000 x 1250
[e] 3000 x 1500
3000 x 1750
[f] 3000 x 2000
4500 x
3000