GKP-2302 User Manual
60
Figure 4-3 Stop Phase
--------
Notes
----------------------------------------------------------------------------------------------------------------
When the Step Behavior is set to [Sweep], output is constant while waiting for the step
Stop Phase after the Step Time passes.
Step Stop Phase setting can be disabled. In this case, when the set Step Time has passed,
the execution transitions to the next step without waiting for step Stop Phase.
In DC mode, the step Stop Phase cannot be set.
---------------------------------------------------------------------------------------------------------------------------------
Trigger output (Trig Out)
State output to CONTROL I/O connector. This specifies whether the trigger output presents
when the step starts. Polarity and pulse width of the trigger output complies with the trigger
output setting (see
4.2.3
Example of Output Using Sequence Function
As an example, consider the sequence that gradually changes DC voltage like Figure 4-4.
Set the output mode to DC-INT. Edit the sequence like Table 4-3. Parameters not included in
this table should be left as the factory default setting.
O
u
tp
u
t
v
o
lt
a
g
e
Time
Start operation
Stop operation
Figure 4-4 Example of Sequence
Table 4-3 Example of Editing Sequence
Step Number
0
1
2
3
Step-contro
l parameter
Step Time (Time)
5 s
20 s
10 s
15 s
Jump-to (Jump To)
—
OFF
OFF
OFF
Step termination (Step
Term)
—
Continue
Continue
Hold
Intra-Step
Parameter
DC Voltage (DCV)
0 V
+100 V
(Optional)
+50 V
Step Behavior
Sweep
Sweep
Keep
Const
--------
Notes
----------------------------------------------------------------------------------------------------------------
Step Time and the Step Behavior of step No. 0 specifies the action after the transition to
step No. 0 at the end of a sequence.
Jump-to of step No. 0 cannot be specified. The execution always transitions to step No. 1
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......