GKP-2302 User Manual
120
Opening the Memory View
1.
Press the MENU key to move to the root menu, then select [Memory] (or press the
MEMORY key on the Continuous function screen).
2.
The Memory View opens.
4.8.1
Basic Setting Memory
The Basic Setting Memory collectively stores the output-related settings in the Continuous
function (e.g. AC/DC mode, signal source, output range, AC setting, DC setting, current limiter,
setting range limit). User is required to perform some operations to save these settings into the
Basic Setting Memory. Both the internal and USB memories can be selected as the Basic Setting
Memory.
There are 31 internal Basic Setting Memories from No. 0 to No. 30. The memory No. 0 stores the
factory defaults. The internal memory No. 1 settings are loaded at power-on of this product.
User can save the settings to the memories from No. 1 to No. 30. Clearing a Basic Setting
Memory resets it to the factory defaults (same as the one of No. 0). Clearing a Basic Setting
Memory in the USB memory deletes the applicable file in the USB memory.
For the set content saved in the Basic Setting Memory and factory defaults, see