GKP-2302 User Manual
166
3.
Enable/Disable the energization. The power unit on which the energization is
enabled/disabled is represented by
/
respectively. Move the cursor to
/
and press
the ENTER key to switch between
and
.
4.
Select [OK]. The energization is enabled or disabled and the window is closed.
--------
Notes
----------------------------------------------------------------------------------------------------------------
The maximum current, current monitor gain, current measurement full scale, and the
power measurement full scale are set to the values corresponding to the rated output
power which is determined by the number of power units to be energized. However, the
applied current measurement accuracy and the power measurement accuracy are the
values under the condition where all power units are enabled for energization.
If the number of the energized power units is changed, the setting values of the peak
current limiter and the RMS current limiter are restored to the factory default settings
corresponding to the rated power which is determined by the number of the power units
to be energized.
If you select [OK] without changing the energization settings in the power unit
energization setting window, the energization settings of all power units are disabled once
and enabled again. The setting values of the peak current limiter and the RMS current
limiter are restored to the factory default settings corresponding to the rated power which
is determined by the number of the power units to be energized.
If a failure occurs in a power unit for some reason, you can disable the energization for
that unit to restart using this product with the remaining power units. For details, see
---------------------------------------------------------------------------------------------------------------------------------
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......