GKP-2302 User Manual
58
DC voltage (DCV)
The output settings of each step. For the phase angle, set the phase angle as the difference from
the L1 phase. Phase angle can also be set per step. The phase angle of L1 phase cannot be set,
and is fixed as 0 deg.
Step Time (Step Time)
Set the output time of that step. Unit can be selected from [s] or [ms].
Jump-to step (Jump To)
When selected,
“to jump” step to
can be specified. By specifying the Jump Count ([Jump Cnt]),
you can set up a loop to repeat the same step a number of times. After jumping for the number
of times specified for one sequence, the action at the end of step follows the step termination
setting.
If Jump-to is set to off, the action at the end of step follows the step termination setting.
--------
Notes
----------------------------------------------------------------------------------------------------------------
When the step termination is set as [Hold], the hold status is enabled before the jump.
---------------------------------------------------------------------------------------------------------------------------------
Jump count (Jump Cnt)
Sets the number
of times to “
jump to
”
the step specified in the Jump-to setting. This can be set
when Jump-to ([Jump To]) is selected. When the Jump count is set to 0, the Jump count is set to
infinity.
--------
Notes
----------------------------------------------------------------------------------------------------------------
The number of times that a particular step is repeated is Jump count +1.
---------------------------------------------------------------------------------------------------------------------------------
Branch step (Branch1, Branch2)
When the step is under execution or in the hold state, the step to transition when the branching
instruction through panel operation or external control is detected can be specified. There are
two systems of branch instructions (Branch1 and Branch2), and an instruction can be set for
each.
--------
Notes
----------------------------------------------------------------------------------------------------------------
Transition using branch is executed immediately after a branching instruction is detected.
Step Time, Stop Phase, and step termination setting are ignored.
---------------------------------------------------------------------------------------------------------------------------------
Step termination (Step Term)
Specifies the action when the step completes. When [Continue] is selected, transitions to the
step of the next number. When [End] is selected, the sequence is terminated and the execution
transitions to step No. 0 and enter into the stand-by state. When [Hold] is selected, the output at
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......