4.4 Measuring Harmonic Current
107
電圧
電流
Voltage
Current
Figure 4-20 Current Waveform Containing Many Harmonic Components
4.4.2
Basics
Harmonic current measurement function is available only for the Continuous function, AC-INT,
and set frequency of 50 Hz or 60 Hz.
The harmonic current components up to order 40 are displayed as the RMS and a percentage to
fundamental wave component.
--------
Notes
----------------------------------------------------------------------------------------------------------------
The Harmonic current measurement function does not support the standard tests defined
by organizations such as IEC. Use this function for a preliminary testing.
---------------------------------------------------------------------------------------------------------------------------------
4.4.3
Measured Value Display Method
Operation procedure
1.
In the Continuous function, select the [Measure] soft-key
[4: Harmonic].