9.18 Sequence
221
9.18
Sequence
Effective only for AC-INT, AC+DC-INT, and DC-INT.
GKP-2303
Number of memories
5 (non-volatile)
Number of steps
Maximum 255 (for each sequence)
Setting range of Step
Time
0.0010 s to 999.9999 s
Intra-Step behavior
Constant, Keep, Linear Sweep
Parameter
Output range
AC/DC mode (The signal source is fixed to INT)
(The above 2 items are common within one sequence)
AC phase voltage, frequency, waveform
DC voltage
Start Phase
Stop Phase
Step termination
Jump count (1 to 9999, or infinite)
Specification of the Jump-to step
Synchronous step output (2bit)
Specification of the branch step
Trigger output
Sequence control
Start
Stop
Hold
Resume
Branch 1, Branch 2
Note: For DC-INT, the AC phase voltage, frequency, waveform, Start Phase, and Stop Phase
cannot be set.
9.19
Simulation
This allows you to simulate power line abnormalities, such as blackout, voltage rise, voltage
drop, abrupt phase change, and abrupt frequency change. It can be used only for AC, sine wave,
and AC+DC-INT. Note that it does not support IEC or other standard test.
GKP-2303
Number of memories
5 (non-volatile)
Number of steps
6 (Initial, Normal 1, Trans 1, Abnormal, Trans 2, Normal 2)
Setting range of Step
Time
0.0010 s to 999.9999 s (0 s is available only for the Transition Step)
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......