Tollerance Description
DIATERMO
MB250
DIATERMO
MB400
15%
Maximum output voltage CUT (V
pp
) 1500
1500
15%
Maximum output voltage CUT/COAG 80% (V
pp
) 1500
1500
15%
Maximum output voltage CUT/COAG 60% (V
pp
) 1500
1500
15%
Maximum output voltage CUT/COAG 40% (V
pp
) 1500
1500
15%
Maximum output voltage ENHANCED CUT (V
pp
) 1500
1500
15%
Maximum output voltage BLEND (V
pp
) 2500
2500
15%
Maximum output voltage SPEEDY COAG (V
pp
) 2500
2500
15%
Maximum output voltage DEEP COAG (V
pp
) 800
800
15%
Maximum output voltage SPRAY COAG (V
pp
) 4500
4500
15%
Maximum output voltage SPRAY COAG ARGON (V
pp
) 4500
4500
15%
Maximum output voltage BIPOLAR CUT (V
pp
) 800
800
15%
Max output voltage BIPOLAR CUT/COAG 80%(V
pp
) 800
800
15%
Max output voltage BIPOLAR CUT/COAG 60%(V
pp
) 800
800
15%
Max output voltage BIPOLAR CUT/COAG 40%(V
pp
) 800
800
15%
Maximum output voltage BIPOLAR BLEND (V
pp
) 1100
1100
15%
Maximum output voltage BIPOLAR COAG (V
pp
) 400
400
0.5
Size LxHxP mm
470x150x380
470x150x380
10
Weight (kg)
18 18
5%
Selectable mains power (V
~
)
115 – 230
115 – 230
1%
Mains frequency (Hz)
50-60
50-60
0
Fuses 230V
~
(5x20) TIMED
2xT 4A
2xT 5A
0
Fuses 115V
~
(5x20) TIMED
2xT 8A
2xT 10A
10%
Electrical input power (VA)
800
1000
10%
Electrical input current (230V
~
) (A)
3.5
4.5
10%
Electrical input current (115V
~
) (A)
7
9
5
Five steps adjustable sound level
Self-check
Power accuracy output warning
Skin Plate Electronic Control
1
1
1
Split or not split patient plate allowed
Working condition storing
2
10
2
10
2
Electrical Class (EN60601-1)
I CF
I CF
MDD 93/42/EEC Class
II b
II b
EN55011 (CISPR 11) Class (Class/Group)
2 / B
2 / B
Patient circuit
-
F
-
-
F
-
Duty Cycle (action / pause) in seconds
10 / 30
10 / 30
Output power control by foot-switch or finger-switch
Defibrillation-proof
Ten seconds delivery warning (OVT)
Equipotential binding
Metallic cabinet
Polycarbonate covered panels
Conform to EN60601-1 (1998)
Conform to EN60601-1-2 (2003)
Conform to EN60601-2-2 (2000)
= STANDARD
1
Patient to plate contact monitoring system
2
Continuous storing of the last settings