43
know the impedance value on the patient, by pressing one time (in not delivery phase) the OK key. The value checks on the
patient is visualized for 2 seconds on the display that, than switches off itself.
7.3.3
P
ROGRAM
In the delivery phase, on the LCD display of the PROGRAM section, the operator receives the informations about used
parameters.
In the case of not delivery, the operator can access to the MENU’ function, visualized on the LCD display of PROGRAM
section, by pressing the acceptance key (enter
)
‘
↵
’ and choice, by skiming through the menù with ‘
↓
’ and ‘
↑
’ keys, between
the four following settings:
1. Save
. Press the acceptance key (enter) ‘
↵
’
to enter in this section in which modify the name (max 11 letters long) of the
program. By the keys (down) ‘
↓
’ and (up) ‘
↑
’ select the letters and save them, once at a time, by the acceptance key
(enter) ‘
↵
’. When the memorization is finished, after the last pressure of the enter key ‘
↵
’ you can esc from this section. If
you don’t want to save the name of the program and to esc from this section, press the esc key ‘
←
’
.
2. Program
. Press the enter key ‘
↵
’ to enter in this section and read the different memorized programs by the keys (down)
‘
↓
’ and (up) ‘
↑
’.
Press the enter key ‘
↵
’ to select the desired program. By pressing the esc key ‘
←
’
you can esc from this
function but without any selection .
3. Errors
. Press the enter key ‘
↵
’ to enter in this section and by the keys
(down) ‘
↓
’ and (up) ‘
↑
’ read the errors list
(
event-error
description is visualized on the seven segments LCD display) verified in the equipment since the last event-
error to the older (more than one hundred memorized errors). To esc from this section press the esc key ‘
←
’.
4. Delay Clamp
. Press the enter key ‘
↵
’ to enter in this section and to set, by the keys (down) ‘
↓
’ and (up) ‘
↑
’, a delay for
the delivery respect to the contact bipolar forceps-tissue. Select the enter key ‘
↵
’ to confirm the set delay. This function
can be set if you have been chosen with automatic bipolar coagulation (see
Autostart and Autostop
paragraph).
7.4 MONOPOLAR
The supplying currents in the monopolar way for cut, coagulated cut and coagulation can be predisposed by the keys present in
the MONOPOLAR 1, MONOPOLAR 2 sections. The power level for every function can be predisposed by the + and – keys
of CUT , COAG1 and COAG 2 sections. The set power levels remain in the memory.
(Fig. 1)
MONOPOLAR 1
CUT COAGULATION
HANDLE
CONNECTOR
1 Cut and coagulated cut (CUT)
4 Speedy Coag
2 Enhanced Cut
5 Deep Coag
3 Blend
Using the CUT or COAG 1 functions it will need to connect the handle to the corresponding output CUT/COAG 1, in the
MONOPOLAR 1 section.