![Gima DIATERMO MB 250 Instruction Manual Download Page 43](http://html1.mh-extra.com/html/gima/diatermo-mb-250/diatermo-mb-250_instruction-manual_2228032043.webp)
GIMA SpA
DIATERMO MB250 – DIATERMO MB400
MA303IGBa-19-03-2007 Manuale
d’Istruzioni / Instruction’s Manual
41
7.
CONNECTORS AND CONTROLS
7.1 LABEL ON THE REAR PANEL
The requirements for the safety of H.F. surgical equipment ask technical data and graphic symbols must be printed on the
cabinet or on at least one of the panels of generator unit to define its features and oversee its condition of work.
7.1.1
M
ANUFACTURERS
’
I
DENTIFICATION
D
ATA
DIATERMO MB250
or
DIATERMO MB400
electrosurgical unit is designed, manufactured and tested by the LED SpA in
its own laboratories in Aprilia (LT) - Italy.
7.1.2
T
ECHNICAL
D
ATA
DIATERMO MB250
MONOPOLAR APPLICATION
BIPOLAR APPLICATION
FREQUENCY:
425kHz FREQUENCY:
525kHz
Output CUT (CUT):
250W - 300
Ω
Output BIPOLAR CUT (CUT):
120W - 150
Ω
Output ENHANCED CUT (CUT):
150W - 350
Ω
Output BIPOLAR BLEND (CUT):
100W - 300
Ω
Output BLEND (CUT /COAG):
150W - 500
Ω
Output BIPOLAR COAG
70W - 100
Ω
Output SPEEDY (COAG.):
100W - 500
Ω
Output DEEP (COAG.):
100W - 200
Ω
Output SPRAY (COAG.):
60W - 2000
Ω
MAIN POWER:
115/230 V - 50/60 Hz selecting
INLET
POWER:
800VA
FUSE:
(230V
~
) 2xT 4A / (115V
~
) 2xT 8A
DUTY - CYCLE:
intermittent 10 sec emission / 30 sec of pause
CLASS:
I
CF
DIATERMO MB400
MONOPOLAR APPLICATION
BIPOLAR APPLICATION
FREQUENCY:
425kHz FREQUENCY:
525 kHz
Output CUT (CUT):
400W - 250
Ω
Output BIPOLAR CUT (CUT):
120W - 150
Ω
Output ENHANCED CUT (CUT):
250W - 250
Ω
Output BIPOLAR BLEND (CUT):
100W - 300
Ω
Output BLEND (CUT /COAG):
250W - 500
Ω
Output BIPOLAR COAG (COAG):
70W - 100
Ω
Output SPEEDY (COAGULATION):
120W - 500
Ω
Output DEEP (COAG.):
120W - 200
Ω
Output SPRAY (COAGL):
100W - 2000
Ω
MAIN POWER:
115/230 V - 50/60 Hz selecting
INLET
POWER:
1000VA
FUSE:
(230V
~
) 2xT5 A / (115V
~
) 2xT10 A
DUTY - CYCLE:
intermittent 10 seconds emission / 30 seconds of pause
CLASS:
I
CF
7.1.3
M
EANING OF
G
RAPHIC
S
YMBOLS
The meaning of the graphic symbols printed on
DIATERMO MB250
or
DIATERMO MB400
’s cabinet is the following:
1. Floating Patient's plate: neither at low-frequency nor at high frequency earth connected
2. The equipment is protected against Cardiac Defibrillator discharge.
3. Not Ionising Radiation emitted
4 Corresponding to the Directive of the Europe Community 93/42/CEE
5. Read carefully INSTRUCTION’S MANUAL before to attempt the use of the equipment.
6- The product mustn’t be threw in the containers for urban wastes but it must be swallowed by a separate picking
.
1
2
3
4
5
6