GIMA SpA
DIATERMO MB250 – DIATERMO MB400
MA303IGBa-19-03-2007 Manuale
d’Istruzioni / Instruction’s Manual
48
Handle connector for the monopolar2 cut and coagulation
This is the point of connection for the handle with double push-buttons (if the
handle without push-buttons, it must be connected to the ‘active’ point) to realize
the functions of the cut CUT and coagulation COAG2.
Remember that with the yellow push-button there is the control for the delivery of
the monopolar currents for the cut and with the blue push-button there is the
control for the monopolar coagulation.
Foot-switch connector for the monopolar2cut and coagulation
This is the point of connection for the foot-switch to realize the monopolar cut and
the coagulation (MONOPOLAR 2).
Remember that by the pressure of the yellow foot-switch the cut function is
activated, while by the pressure of the blue one the coagulation function is
activated.
Bipolar forceps connector
Point of connection for the bipolar forceps, by which the bipolar currents can be
distributed to realize the bipolar cut and coagulation.
Double foot-switch for the bipolar cut and coagulation
Point of the foot-switch connection.
Remember that by the pressure of the yellow
foot-switch the cut function is activated, while by the pressure of the blue one the
coagulation function is activated.
7.11 BACK PANEL
1 Spendthrift
6 Connector for external unit
2 Socket RS-232 (for assistance services)
7 Mains mechanical switch
3 Socket RS-232 (for assistance services)
8 Mains socket
4 Speaker Volume
9 Socket for the equipotential connection
5 Connector for external Argon unit
10 Fuse holder / Voltage selector
7.11.1
C
ONNECTORS
1 - Connector for external Argon unit
Point of connection to an external Argon unit.
After the push-button for the selection of the spray
Argon coagulation has been pressed and by
activating the delivery (preferably by the foot-
switch with the handle without push-buttons), the
connector 1 received a signal that activate the
delivery of this gas.
2 – Connector for external unit (except Argon
unit)
Point of connection to an external unit, except
Argon unit) (see point 1) to which the signal of
delivery comes or, for example, of activation of
surgical aspirator.