52
9.
MAINTENANCE
9.1 GENERAL
No user adjustable parts are within the equipment, either for calibration or service purposes. The equipment housing must not
be opened: the warranty is invalidated by any unauthorized entry into the unit. In the event any repair or adjustment work
being necessary, the whole equipment should be returned to the LED S.p.A. Service Centre 04010 APRILIA (LT) - ITALY, or
to a other Authorised Centre, together with a description of the fault.
Maintenance work by the user is mainly the cleaning of the exterior of the cabinet, cleaning and sterilisation of the accessory
items and checking of the equipment before each use. Carrying out function and safety check for verification of the parameters
is demanded to specialised technical people.
9.2 CLEANING OF THE CABINET
Switch the equipment off completely and disconnect the mains supply before any cleaning is undertaken. Clean the outside of
the cabinet with a damp cloth. No chemical should be used; a mild non abrasive cleanser may be used when necessary.
9.3 CLEANING AND STERILISATION OF THE ACCESSORY ITEMS
The best thing to do is to use only one time use accessories and discard them after use. Since some of the accessory items are
to be used more than once it is mandatory to clean carefully and sterilize those accessories before the new use. The best way to
clean and sterilize the reusable items is to follow the direction of the supplier of each item. When original reusable accessories
supplied by LED S.p.A. are applied, the cleaning by using deep cleanser and sterilization through steam sterilization at 121 °C
is recommended.
9.4 GUIDE TO THE SOLUTION OF THE PROBLEMS
In case of problems before all it is advised to check for the correct installation of the unit and for the correct connection of the
accessories. The code of the produced error is visualized on the seven segments display, while the description is on the LCD
display.
Problems
Probable Cause
Solution
The equipment doesn't
switch on.
Interruption or absence of the main feeding Verify the connection of the main cable.
Verify the fuses and replace them, where necessary, with new
ones of the propre type.
Alarm OC always active.
Interruption or lack of contact on the
neutral electrode circuit
Check the connection of the cable to the neutral electrode.
Replace the cable of connection of the neutral electrode.
The unit doesn't respond to
the command of activation.
Breakdown of the handpiece or of the pedal
- Wrong connection of the handpiece or of
the pedal - Alarm OVT activated
Replace the handpiece or the pedal.
Verify the connection of the handpiece or of the pedal.
Wait for the OVT warning signal getting out.
Error Code 001
Current delivery control activated during
switching on
Disconnect the handpiece or the pedal and switch on the unit
again.
Error Code 002
Error in the management board
Call for Service
Error Code 003
Error in the management board
Call for Service
Error Code 004
Error in the data conversion circuit
Call for Service
Error Code 005
Error of the reference voltage value
Verify the main voltage
Call for Service
Error Code 009
Error in the output power activation circuit Call for Service
Error Code 011
Foot switch pressed
Verify the state of the foot-switch
Error Code 013
Bipolar DAC not verified
Call for Service
Error Code 014
Bipolar Power not verified
Call for Service
Error Code 016
Fuse module blown BIPO
Call for Service
Error Code 017
Fuse blown 12V o -8V BIPO
Call for Service
Error Code 018
Fuse blown 20 BIPO
Call for Service
Error Code 019
Fuse blown +HV BIPO
Call for Service
Error Code 020
MONO. broken circuit
Call for Service
Error Code 021
BIPO. broken circuit
Call for Service
Error Code 022
Info MONO interrupted
Call for Service
Error Code 023
Info BIPO interrupted
Call for Service