![Gima DIATERMO MB 250 Instruction Manual Download Page 61](http://html1.mh-extra.com/html/gima/diatermo-mb-250/diatermo-mb-250_instruction-manual_2228032061.webp)
GIMA SpA
DIATERMO MB250 – DIATERMO MB400
MA303IGBa-19-03-2007 Manuale
d’Istruzioni / Instruction’s Manual
59
DIATERMO MB400
CUT100% - Vpeak max (on 5200 Ohm)
0
100
200
300
400
500
600
700
800
900
1000
0
40
80
120
160
200
240
280
320
360
400
Setting
Vp
e
a
k
Diagrams of CUT100% maximum mains voltage output versus Vp
CUT80% - Vpeak max (on 5200 Ohm)
0
100
200
300
400
500
600
700
800
900
1000
0
30
60
90
120
150
180
210
240
270
300
Setting
Vp
e
a
k
Diagrams of CUT80% maximum mains voltage output versus Vp
CUT60% - Vpeak max (on 5200 Ohm)
0
100
200
300
400
500
600
700
800
900
1000
0
25
50
75
100
125
150
175
200
225
250
Setting
Vp
e
a
k
Diagrams of CUT60% maximum mains voltage output versus Vp
CUT40% - Vpeak max (on 5200 Ohm)
0
100
200
300
400
500
600
700
800
900
1000
0
20
40
60
80
100
120
140
160
180
200
Setting
Vp
e
a
k
Diagrams of CUT40%maximum mains voltage output versus Vp
ENHANCED - Vpeak max (on 5200 Ohm)
0
100
200
300
400
500
600
700
800
900
1000
0
25
50
75
100
125
150
175
200
225
250
Setting
Vp
e
a
k
Diagrams of ENHANCED maximum mains voltage output versus
Vp
BLEND - Vpeak max (on 5200 Ohm)
0
150
300
450
600
750
900
1050
1200
1350
1500
0
25
50
75
100
125
150
175
200
225
250
Setting
Vp
e
a
k
Diagrams of BLEND maximum mains voltage output versus V p
SPEEDY COAG - Vpeak max (on 5200 Ohm)
0
150
300
450
600
750
900
1050
1200
1350
1500
0
20
40
60
80
100
120
Setting
Vp
e
a
k
Diagrams of SPEEDY COAG maximum mains voltage output
versus Vp
DEEP COAG - Vpeak max(on 5200 Ohm)
0
50
100
150
200
250
300
350
400
450
500
0
20
40
60
80
100
120
Setting
Vp
e
a
k
Diagrams of DEEP COAG maximum mains voltage output versus
Vp
SPRAY COAG - Vpeak max (on 5200 Ohm)
0
300
600
900
1200
1500
1800
2100
2400
2700
3000
0
20
40
60
80
100
Setting
Vp
e
a
k
Diagrams of SPRAY COAG maximum mains voltage output versus
Vp
BIPOLAR CUT100% - Vpeak max (on 5200 Ohm)
0
50
100
150
200
250
300
350
400
450
500
0
20
40
60
80
100
120
Setting
Vp
e
a
k
Diagrams of BIPOLAR CUT 100% maximum mains voltage output
versus Vp
BIPOLAR CUT80% - Vpeak max (on 5200 Ohm)
0
100
200
300
400
500
0
20
40
60
80
100
Setting
Vp
e
a
k
Diagrams of BIPOLAR CUT 80% maximum mains voltage output
versus Vp
BIPOLAR CUT60% - Vpeak max (on 5200 Ohm)
0
50
100
150
200
250
300
350
400
450
500
0
20
40
60
80
100
Setting
Vp
e
a
k
Diagrams of BIPOLAR CUT60% maximum mains voltage output
versus Vp
BIPOLAR CUT40% - Vpeak max (on 5200 Ohm)
0
50
100
150
200
250
300
350
400
450
500
0
10
20
30
40
50
60
Setting
Vp
e
a
k
Diagrams of BIPOLAR CUT40% maximum mains voltage output
versus Vp
BIPOLAR BLEND - Vpeak max (on 5200 Ohm)
0
75
150
225
300
375
450
525
600
675
750
0
20
40
60
80
100
Setting
Vp
e
a
k
Diagrams of BIPOLAR BLEND maximum mains voltage output
versus Vp
BIPOLAR COAG - Vpeak max (on 5200 Ohm)
0
20
40
60
80
100
120
140
160
180
200
220
240
0
10
20
30
40
50
60
70
Setting
Vp
e
a
k
Diagrams of BIPOLAR COAG maximum mains voltage output
versus Vp
Information about elimination of this product (Applicable in the European
Union and other European countries with separate collection systems)
On the end of the life, the present product mustn’t be
eliminated as urban refusal, but it must be eliminated in a
separated collection.
If the product is eliminated in unsuitable way, it is possible
that some parts of the product (for example some
accumulators) could be negative for the environment and for
the human health.
The symbol on the side (barred dustbin on wheel) denotes
that the products mustn’t throw into urban refuses container
but it must be eliminated with separate collection.
In case of abusive elimination of this product, could be
foreseen sanctions.