94
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
B.3 VLF OPTION
If you select
A-all
from the
Send Menu
(i.e. to transfer all VLF data), the format is as follows:
141645.0 0000N 0000100E 57137.66 57001.37 000N 21.4 006.0 000.9 071 -026 005.6
24.0 -013.4 001.1 -000 090 012.1 23.4 -027.7 -003.1 -004 066 000.5
141645.0
Time, as hhmmss.s
0000N
Line number and direction
0000100E
Station number and direction
57137.66
Uncorrected magnetic field (nT)
57001.37
Corrected magnetic field (nT)
000N
Slope (degrees)
21.4
First VLF station frequency
006.0 ip - vertical in phase component (%)
000.9 op - vertical out of phase component (%)
071
x - horizontal amplitude
-026
y - horizontal amplitude
005.6
VLF total field strength (pT)
24.0
Second VLF station frequency
-013.4 ip
(%)
001.1
op (%)
-000
x amplitude
090
y
amplitude
012.1 VLF total field strength (pT)
The next six numbers represent the same values for the third VLF station.