5.4 WORKING WITH MOBILE MODE (ALL MODELS)
As described in the instrument description section in Chapter 4, GEM’s ROVER modes include
mobile, gradiometer, walking and walking gradiometer. This section describes the mobile mode that
(along with the BASE STATION mode) is a standard feature of all GSM-19 series units.
5.4.1 Performing a Reading
Mobile mode is a standard mode and is always included in every GSM-19 configuration.
CAUTION:
Before taking a reading, check to make sure all cables and sensor(s) are
connected. This maximizes the lifespan of your system and protects internal RF
circuits from water damage, overheating, etc.
With
MOBILE
survey mode selected, press
A-start
from anywhere in the
Survey Menu
.
SCREEN 33
survey mode
position time file
cycle time tuning AC filter
display mode text ID
connect sensor now
029127 readings left
mobile
A-start C-change
←
BF
→
•
If tune initialize is selected (see section 5.3.7), the unit displays
Please wait
while scanning
through its range for the initial tuning value.
•
If
text
is selected in
Display Mode
(see section 5.3.9) the following screen is displayed:
SCREEN 34
A - menu 1 - repeat (same position)
other keys - read
L 100 N S 200 E
To take a reading, press any other key (with the exception of
A
which returns you to the menu).