72
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
SCREEN 67
50 110 300
600 1200 24000 48000
9600 14400 19200 28800 38400
57600
152000
A-select
←
BF
→
Scroll backward with
B
or forward with
F
and highlight the desired transmission rate (keep pressed
for a continuous scroll).
Press
A-select
to save your option. The unit returns to screen 66.
Pressing
F-ok
leads you to the selection of RS-232 parameters for Real-time Transmission (RTT).
When RTT is enabled, the unit outputs the readings (tagged with their time stamp) to the RS-232 6
pin connector in real time as they are taken.
SCREEN 68
real time RS-232 transmission
yes
F-ok C-change
C-change
toggles between yes and no.
•
If you do not intend to use the RTT function, select
no
and press
F-ok
to return to the
Info
Menu.
Then press
1C
to return to
main menu
.
•
Otherwise select
yes
and press
F-ok
. The
RTT
setting screen is displayed.