56
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
Select
C-grad.
The following screen is displayed. This screen enables you to define settings for the
gradiometer mode.
SCREEN 46
survey mode
position time file
cycle time tuning AC filter
display mode text ID sensors
connect sensors now
029127 readings left
grad
A-start C-change
←
BF
→
5.5.2 Specifying Gradiometer Sensor Spacing
Press
F
to highlight
sensors
. The following screen is displayed.
SCREEN 47
survey mode position time file
cycle time tuning AC filter
display mode text ID
sensors
56 cm apart
A-start
C - change
←
BF
→
NOTE:
This screen is available only with the gradiometer models.
C-change
allows you to set the distance between sensors. The change is allowed only once per file
before the first reading is taken.
Grad displayed is in nT/m and is obtained as
Gradient (nT/m) = FIELD lower channel (nT)
-
FIELD upper channel (nT)
Distance between sensors (m)
If just the difference of the two fields is required, enter 100cm as the distance between the sensors.