65
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
•
If none of the units is equipped with GPS, set the time manually in one of them
and use the synch cable to transfer the time to the other. In this case, both
units run on their respective internal timers.
5.8.2 Transferring the Time
To transfer the time, from the
Main Menu
press
E-time-synch
on EACH unit. The following
screen is displayed.
SCREEN 57
connect cable
& press C
Connect the 6-pin to 6-pin cable between the two units.
NOTE:
The same cable is used for diurnal correction. This is why sometimes it is called
Reduction Cable and sometimes Synch cable. Please contact GEM Systems if you
need to order one.
Press
C
.
The following screen is displayed.
SCREEN 58
A-source
B-destination
enter destination first