73
GSM19 Series Magnetometers - Version 7
GEM Systems, Inc. Advanced Magnetometers
For more technical information, visit www.gemsys.ca
SCREEN 69
19200 28800 38400 57600
115200
A-select
←
BF
→
Press
A-select
to save your option. If finished press
F-ok
to return to the
Info Menu
, and then
press
1C
to return to
main menu
.
IMPORTANT NOTES:
The data bits, stop bits and parity are not programmable for either SEND or RTT
but the default values are displayed for your reference.
You MUST set the receiving device to these parameters
8 data bits
1 stop bit
no
parity
The slowest transmission rate for RTT is 19200 bps
If RTT is not enabled, the external triggering of the unit by the RS-232 line may
not work properly or at all since the transmission rate is not initialized (see end
of section 4.3)
6.2.2 Transferring the Data
For proper data transfer, you
MUST
match the RS-232 settings of the magnetometer and those of
the receiving device (PC, data logger, etc).
Remember that you can program the transmission rate for SEND and RTT on the unit. However,
the data bits, stop bits and parity are not programmable for either SEND or RTT, but the default
values are displayed on the setting screens for your reference.