The way the EOL operates depends on the positioning system in use.
a)
EOL with Line / Station system
•
Line Coordinate: The Line Number is incremented by the current Line Increment.
•
Station Coordinate: The Station Number remains the same but the Station Increment reverses
sign.
With this simple process the operator is ready to walk to the next line of the grid and start taking the
readings moving in the opposite direction he or she walked the previous one.
However, if the next line starts at a different station number, you may press
A-
or
B+
to change the
current station number backward or forward by the pre-set station increment.
SCREEN 41
A- B + F-ok
L00100N S0000225E
b)
EOL with X / Y coordinates
•
X
and
Y
coordinates both increment by their respective EOL increment
•
X
and
Y
increments both reverse sign
If you need to make further adjustments, you may press
A-
or
B+
to change the current X
and
Y
coordinates backward or forward by their respective increments.
NOTE:
To walk along X lines set Y increment and X EOL increment both to zero.