Chapter 2: System Description
________________________________
1
Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
InnovOx
On-Line TOC Analyzer
Operating Specifications
1
Analysis Modes
NPOC, TOC (TC-IC), TC, IC
Range
0.250 to 50,000 ppm TOC
Precision
RSD
≤
3% of reading at
>
5 ppm NPOC or TOC
Accuracy
Greater of ±
3% of reading or
±
0.25 ppm, 1 to 100 ppm, NPOC or TOC
Linearity
R
2
≥ 0.995, measured as NPOC
LOD
250 ppb NPOC
1
TOC Calibration Stability
Up to 6 months
Analysis Time
2.6 to 8.3 minutes in Online mode
Particle size
≤
200
μm without filtration or >200 μm with optional wastewater sampler
Sample Temperature
10-60
q
C (41-140
q
F),
,
or 10-85
q
C (41-185
q
F) with high-temperature options
Ambient Temperature
10-40
q
C (50-104
q
F)
Source Sample Pressure
125 psig maximum with provided control valve
0.5 psig minimum without optional sample pump
Minimum Sample Flow Rate
50 mL/min without optional sample pump
Drain
Gravity Drain
Analyzer Specifications
Outputs
External USB Port (1); Internal USB Ports (2); Ethernet (1);
Binary End-of-Analysis Output (1);
Isolated 4-20 mA Analog Outputs (5 analytical; 1 for stream ID)
Inputs
Binary Input (1) for Remote Analysis Start/Stop
Display
Color LCD w/touch-screen
Power
100 – 240
±
10% VAC, 50 – 60 Hz, 400 VA
Alarm Relays
System Fault Alarm (1); User-Configurable Alarms, High and Low Limit (4)
Communications Protocols
Modbus
®
; TCP/IP over Ethernet
Installation/Overvoltage
Category
II
Sample Streams
2 Streams (standard), 5 Streams (optional)
Dimensions
H: 92.7 cm (36.5 in.); W: 64.6 cm (25.4 in.); D:38.7 cm (15.3 in)
Weight
36.3 kg (80 lb)
Safety Certifications
CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2 NO. 61010-1
Environment
Enclosure Options
IP56/NEMA 4X/Class 1 Division 2/ATEX
Maximum Relative Humidity
Up to 95%, non-condensing
Maximum Altitude
3,000 m (9,800 ft)
Pollution Degree
2