Chapter 6: Analyzer Operation
To set a daily flush start time
1. Before you begin, ensure that you have turned
On
the Daily Flush, as directed in the previous section.
2. On the
Configure Online
screen, press the
Daily Flush Start
button. A numerical keypad appears to the
right.
3. Enter the start time in an HH:MM, 24-hour format.
Summary
You can use the
Configure Online
screen (
Setup tab
Online button
Configure Online
) to display the current
configuration for all streams and the check standard, press the
Summary
button. The summary is a convenient
way to verify the configuration before starting the analysis.
Grab Measurements
Use the Grab option to specify settings for Grab analysis when performing analysis on samples contained in 40-
mL vials placed in the Vial Sampler.
To specify Grab analysis settings
1. On the
Setup
tab, press the
Grab
button. The
Analysis Protocol
button appears on the right.
2. Press the
Analysis Protocol
button. The
Protocol
screen appears with a listing of protocols.
Figure 50: The Protocol Screen